Human Nails Metabolite Analysis: A Rapid and Simple Method for Quantification of Uric Acid in Human Fingernail by High-Performance Liquid Chromatography with UV-detection.

Xi-Ling Li,Gao Li,Ying-Zi Jiang,Dongzhou Kang,Cheng Hua Jin,Qing Shi,Toufeng Jin,Koichi Inoue,Kenichiro Todoroki,Toshimasa Toyo'oka,Jun Zhe Min
DOI: https://doi.org/10.1016/j.jchromb.2015.08.044
IF: 3.318
2015-01-01
Journal of Chromatography B
Abstract:A rapid and simple analytical method for the quantification of uric acid (UA) in human fingernails by high-performance liquid chromatography (HPLC) with ultraviolet (UV) detection is described. UA was extracted from human fingernail samples at 90°C for 20min, then separated on an Inertsil ODS-2 column (250×4.6mm I.D., 5.0μm, GL Sciences) by isocratic elution using methanol: 74mM phosphate buffer (pH 2.2) 2:98 (v/v). An UV detector was used to monitor at 284nm. The results indicated that under optimized measurement conditions results were achieved within 8.0min, and a good linearity was achieved from the calibration curves (r(2)>0.9999) in the range of 1.0-10000ng; the limit of detection (S/N=3) was 2.0pg, the inter-day and intra-day assay precisions were all less than 0.46% and the mean recoveries (%) of the uric acid spiked in the human fingernail were 101.95%. The amounts of UA in the fingernails of healthy volunteers were determined.
What problem does this paper attempt to address?