Sample storage time and temperature on the analysis results of Beckman Coulter STKS automated hematology analyzer

林军,黄宇烽,刘玉秀,徐建平,孟泽
DOI: https://doi.org/10.3969/j.issn.1008-8199.2002.01.009
2002-01-01
Abstract:Objectives:To observe the effect of storage time and temperature of samples on the results determined by automated hematology analyzer Beckman Coulter STKS. Methods:Blood from 10 health volunteers were drawn into anticoagulant tubes and divided into three parts, every part included 10 tubes, samples were kept at 4℃,22℃ and 37℃, the whole blood cells were determined by automated hematology analyzer Beckman Coulter STKS in five different time of 0,5,8,12 h and 24 h, and the results of the 3 parts were compared. Results: When samples were kept at 4℃,results of WBC,HGB and MCV were steady in 24h( P 0.05 ). But RBC and PLT were different statistically after 8 h ( P 0.05); When samples were kept at 22℃, WBC, RBC and HGB were not different statistically within 12 h( P 0.05), PLT and MCV were different after 12 h and 5 h ( P 0.05);When kept at 37℃,only HGB was stable in 24 h: WBC and RBC changed significantly after 12 h( P 0.05) and PLT and MCV changed statistically after 5 h( P 0.05). The three group white blood cell scatterplot of the three different temperature after 24 h were abnormal. Conclusions: For automated hematology analyzer Beckman Coulter STKS, the best temperature for sample storage was 4℃ ;and the samples should be determined within 5 h in the ambient temperature.
What problem does this paper attempt to address?