Quantitative Determination for Nerve Growth Factor Concentration in Commercial Product by SEC‐HPLC

Da Huang,Jianrong Zhang,Yongchao Fu,Liming Lu,Yiming Xu,Lingyan Ma,Minxiang Li,Lingyuan Xiong,Hongwei Ren
DOI: https://doi.org/10.5376/bm.2011.02.0001
2011-01-01
Bioscience Methods
Abstract:The commercial product, Mouse Nerve Growth Factor (NGF) for Injectioni¼Nobex®, is consisted of NGF and human albumin. These two components are both proteins and make it difficult to determine the concentration of NGF by conventional methods. In order to establish a new efficiency quality control process for NGF, this work choose Size Exclusion High Performance Liquid Chromatography (SEC-HPLC) to detect stock solutions and products under screened chromatographic conditions: using Shodex KW-802.5 column, at room temperature with the 0.05 M PBS pH=7 buffer as mobile phase, 0.8 mL/min flow rate and detected under 280 nm wavelength. The standard curves with correlation coefficient of 0.999 82 (for stock solutions) and 0.999 97 (for products) were obtained by external standard method. The NGF concentrations tested in HPLC were compared with which in BCA for the same batches, and the RSD values were less than 3%, which corresponded with China Pharmacopoeia (2010). The results show that SEC-HPLC method is sensitive, accurate and reliable for NGF on-line quality control and has the potential for the quantitative determination of other mixed protein preparations.
What problem does this paper attempt to address?