Experimental Study Of Static And Dynamic Characteristics Of A Miniature 6-Axis Force And Torque Sensor

kun li,bo pan,yili fu,shuguo wang
DOI: https://doi.org/10.1109/ICInfA.2015.7279538
2015-01-01
Abstract:This paper describes an experimental evaluation of a miniature force and torque sensor in robot-assisted minimally invasive surgery (RMIS). A sensor prototype including the mechanical system and sensing electronics is overviewed. A linear square based static decoupling method is introduced, design and use of a newly invented calibration platform is presented. A dynamic loading mechanism to simulate a heartbeat motion is described. The sensing ranges of the sensor prototype is +/- 30 N and +/- 30 N.mm with resolutions of 0.08 N (x, y directions), 0.25 N (z direction), and 2.4 N.mm (rotational directions). It also exhibits a good capability to follow the dynamic loads at a frequency close to a normal heart rate of an adult.