Image Reconstruction from Sparse Data Samples in MRI Accounting for Phase Roll

Samuel J. LaRoque,Emil Y. Sidky,Gregory S. Karczmar,Xiaochuan Pan
DOI: https://doi.org/10.1117/12.769568
2008-01-01
Abstract:In this study we apply a total variation (TV) minimization algorithm to image reconstruction in magnetic resonance imaging (MRI). This algorithm is particularly effective for underlying images that are approximately piecewise constant. While the underlying proton spin density in MRI can satisfy this condition under. certain circumstances, it is often distorted by unavoidable physical factors that alter the phase of the complex image. In this work, we employ a known method of removing this slow phase variation resulting from magnetic field inhomogeneities to obtain a spin density distribution that is piecewise constant. After the phase removal, we apply the TV minimization algorithm to obtain images from 20% of the full MRI data set.
What problem does this paper attempt to address?