Field Test of Microbend Fiber Sensor for Hospital Use

Zhihao Chen,Doreen Lau,Ju Teng Teo,Soon Huat Ng,Xiufeng Yang,Pin Lin Kei
DOI: https://doi.org/10.1117/12.2059417
2014-01-01
Abstract:In this paper, we report a field test of a microbend fiber sensor for simultaneous measurement of breathing rate, breathing pattern, Ballistocardiogram and heart rate during magnetic resonance imaging (MRI). Comparative experiments conducted between our sensor and commercial physiologic device on a healthy male subject showed an accuracy of +/-2bpm for simultaneous measurement of both breathing rate and heart rate. Our preliminary field test on simultaneous measurement of breathing rate and heart rate in a clinical trial conducted on 11 healthy subjects in the 3.0 Tesla MRI environment showed very good agreement compared with measurements obtained from conventional MRcompatible devices.
What problem does this paper attempt to address?