Transcranial Photoacoustic Tomography of the Monkey Brain

Liming Nie,Chao Huang,Zijian Guo,Mark Anastasio,Lihong V. Wang
DOI: https://doi.org/10.1117/12.907060
2012-01-01
Abstract:A photoacoustic tomography (PAT) system using a virtual point ultrasonic transducer was developed for transcranial imaging of monkey brains. The virtual point transducer provided a 10 times greater field-of-view (FOV) than finite-aperture unfocused transducers, which enables large primate imaging. The cerebral cortex of a monkey brain was accurately mapped transcranially, through up to two skulls ranging from 4 to 8 mm in thickness. The mass density and speed of sound distributions of the skull were estimated from adjunct X-ray CT image data and utilized with a time-reversal algorithm to mitigate artifacts in the reconstructed image due to acoustic aberration. The oxygenation saturation (sO(2)) in blood phantoms through a monkey skull was also imaged and quantified, with results consistent with measurements by a gas analyzer. The oxygenation saturation (sO(2)) in blood phantoms through a monkey skull was also imaged and quantified, with results consistent with measurements by a gas analyzer. Our experimental results demonstrate that PAT can overcome the optical and ultrasound attenuation of a relatively thick skull, and the imaging aberration caused by skull can be corrected to a great extent.