Development of Indirect ELISA for the Determination of 2,4-D in Water

YU Ruo-zhen,HE Miao,SHI Han-chang,QIAN Yi
DOI: https://doi.org/10.3969/j.issn.1673-5897.2006.01.015
2006-01-01
Abstract:The indirect competitive ELISA was developed for the determination of herbicide 2,4-D in water sample with selfmade specific monoclonal antibody 6D11 against 2,4-D. The dose - response curves were plotted when coating antigen 2,4-D-BSA concentration respectively were 240ng·mL-1, 120ng·mL-1 and 60ng·mL-1. Lower IC50 value was taken when coating antigen concentration was 60ng·mL-1. The dose-response curves were also plotted when competitive step respectively was within 60min and 15min. Lower IC50 value was taken when the competitive step was within 15min. 2,4-D concentration in PBS, drinking water, ground water of Tsinghua University and surface water of Fuhai Lake in Yuanmingyuan Park were detected using above optimum reaction condition. It was found that the matrix effects interfered in detection precision of water samples. The matrix effects were basically eliminated using PBS buffer of 5% ethanol. 2,4-D concentration of 0.5mg·L-1, 0.125mg·L-1 and 0.03mg·L-1 in water samples were determined at above optimal reaction conditions. The recoveries of 2,4-D varying from 70% to 130%, which is approximately according to the precision standard of trace organic pollutants. But the coefficients of variation (CV) were relatively larger. The established immunoassay should be improved in order to practical determination of 2,4-D in water samples.