Development of a Miniature 2-Dofs Bending Manipulator with Bipolar Coagulator

Hiromasa YAMASHITA,Kiyoshi MATSUMIYA,Ken MASAMUNE,Etsuko KOBAYASHI,Ichiro SAKUMA,Hongen LIAO,Makoto HASHIZUME,Takeyoshi DOHI
DOI: https://doi.org/10.5759/jscas1999.9.91
2007-01-01
Abstract:We developed a miniature 2-DOFs bending manipulator of 5-mm diameter for laparoscopic surgery. Bending directions of the mechanism are perpendicular and the mechanisms are controllable respectively. A new wire-connected linkage-driven bending mechanism is built in the manipulator to drive rotate joints for 2-DOFs motion. This manipulator has a central channel of 1.3-mm diameter to drive a pair of blades with electrodes as the tip-side bipolar-coagulator. Surgeons control the manipulator with a grip-type interface in one hand. In mechanical performance evaluations, bending angle was maximum 153.9degrees (from -71.6 to 82.3 degrees) with positioning reproducibility of maximum 0.7mm. About generated force, holding force of bending mechanism was more than 1.82N, and grasping force by a pair of blades as maximum 3.70N. In vivo experiment using a pig (38kg, male) with laparoscopy, we evaluated whether the manipulator performed as a suitable bipolar coagulator for practical clinical use. We were able to coagulate living tissues and occlude blood vessels on the mesenterium completely. In conclusion, our new bending mechanism is useful for miniaturization of the laparoscopic manipulator with the simple structure and high mechanical performance.
What problem does this paper attempt to address?