Research of the Fluorescence Detection Apparatus for Nutrients

Yu Wang,Huimin Yan,Xuxiang Ni,Xiaoyi Xu,Shibing Chen
DOI: https://doi.org/10.1117/12.2202740
2015-01-01
Abstract:The research of the multifunctional analyzer of Clinical Nutrition, which integrates the absorbance, luminescence, fluorescence and other optical detection methods, can overcome the functional limitations of a single technology on human nutrition analysis, and realize a rapid and accurate analysis of the nutrients. This article focuses on the design of fluorescence detection module that uses a photomultiplier tube(PMT) to detect weak fluorescence, and utilizes the single photon counting method to measure the fluorescence intensity, and then according to the relationship between the fluorescent marker and fluorescence intensity, the concentration of the analyte can be derived. Using fluorescein isothiocyanate(FITC, the most widely used fluorescein currently)to mark antibodies in the experiment, therefore, according to the maximum absorption wavelength and the maximum emission wavelength of the fluorescein isothiocyanate, to select the appropriate filters to set up the optical path. In addition, the fluorescence detection apparatus proposed in this paper uses an aspherical lens with large numerical aperture, in order to improve the capacity of signal acquisition more effectively, and the selective adoption of flexible optical fiber can realize a compact opto-mechanical structure, which is also conducive to the miniaturization of the device. The experimental results show that this apparatus has a high sensitivity, can be used for the detection and analysis of human nutrition.