Estradiol Concentration in Rana Oviduct, Ovum of Eriocheir Sinensis and Ovum of Crucian Carp by High Pressure Liquid Chromatography

Ting LI,Zheng WANG,Hai-Bo ZHANG,Li-Li JIN,Qiu-Yu WANG,Yu-Zhu MA
2014-01-01
Abstract:Objective To determine the concentration of estradiol in Rana oviduct, ovum of Eriocheir si-nensis and ovum of Crucian carp by high pressure liquid chromatography. Methods The chromatographic column Agilent Zorbax SB-C18 (4.6 mm×250 mm, 5 μm) was used. Chromatographic conditions were as fol-lows:the mobile phase was acetonitrile-water (63:37), the detection wavelength was 280 nm, the flow rate was 1.0 mL/min, column temperature was 30 ℃, and sample amount was 10μL. Results A good linearity were obtained in the range of estradiol concentration 0.3~10μg/mL with the above condition. Its regression equation was Y=4203.8X-661.95 with the correlation coefficient R=0.9991. The estradiol concentration of Rana oviduct, ovum of Eriocheir sinensis and ovum of Crucian carp were 3.00, 0.42, 0.54 μg/g, respectively. Conclusion The estradiol in these samples could be extract effectively by ultrasonic method. The optimized method has a good relationship and recovery rate for estradiol, and its peak pattern and separating effect of HPLC was good.