Non-uniformity Calibration and Correction of Reflector-Based Mosaic CCD Camera

LI Xiao-jie,REN Jian-wei,LI Xiang-sheng,SUN Jing-xun,WAN Zhi,LIU Hong-xing
DOI: https://doi.org/10.3788/yjyxs20142906.1057
2014-01-01
Abstract:According to the radiometric calibration experiments in laboratory,the reflector-based mo-saic CCD camera’s image non-uniformity caused by vignetting phenomenon was corrected.Firstly, the image un-uniformity reasons of reflector-based mosaic CCD camera are analyzed,and the irradiance deviation received by different CCD pixels in focal plane is the essential reason for images’non-uni-formity. Then, the CCD pixels ’ irradiance responsivity calibration experiment and the whole camera’s non-uniformity calibration experiment were carried out in the laboratory,and the relation-ship between each pixel’s receiving irradiance and camera’s incident radiance was built.Lastly,a cor-rection algorithm is put forward based on the above relationship to calculate CCD pixels’correction factors,and the correction factors were used to correct the vignetting image.The correct experiment indicates that,within a certain range of incident radiance,the non-uniformity of original calibration images are above 1 6%,and after correction,the values decreased to 1% or less.Besides,the correc-tion factors were used to correct the images of outdoor experiments and it works very well.