Chemiluminescence Determination of Melamine with Luminol-K 3 Fe(CN) 6 System

Xiao-Shuang Tang,Xi-Yan Shi,Yu-Hai Tang,Zhong-Jin Yue,Qi-Qi He
DOI: https://doi.org/10.1016/s2095-1779(11)70018-1
2011-01-01
Abstract:A sensitive chemiluminescence (CL) method was developed for determining melamine in urine and plasma samples based on the fact that melamine can remarkably enhance the chemiluminescence of Luminol-K3 Fe(CN)6 system in alkaline medium. The determination conditions were optimized. Under optimum conditions, the chemiluminescence intensity had a good linear relationship with melamine in the range of 9.0×10−9 – 7.0×10−6g/mL with a correlation coefficient of 0.9992. The detection limits (3σ) were 3.54ng/mL for urine sample and 6.58ng/mL for plasma sample. The average recoveries of melamine were 102.6% for urine sample and 95.1% for plasma sample. Melamine in samples was extracted with liquid-liquid extraction procedures and the assay results coincided very well with that determined with flow injection chemiluminescence method. The method provides a reproducible and stable approach for sensitive detection and quantification of melamine in urine and plasma samples.
What problem does this paper attempt to address?