Robust iterative image reconstruction for breast CT by use of projection differentiation

david n kraemer,erin g roth,emil y sidky,i reiser,xiaochuan pan
DOI: https://doi.org/10.1117/12.2082444
2015-01-01
Abstract:Design of image reconstruction algorithms for breast CT faces challenging conditions: projection data are often truncated particularly near the chest wall, and the noise level in the acquired projections is high when constraining the scan to have equivalent dose as two-view mammography. Recently, we have been investigating a design of iterative image reconstruction algorithms that employ a differentiation filter on the projection data and estimated projections. The extra processing step can potentially reduce the impact of artifacts due to projection truncation in addition to enhancing edges in the reconstructed volumes. The edge enhancement can improve visibility of various tissue structures. Previously, this idea has been incorporated in an approximate solver of the associated optimization problem. In the present work, we employ singular value decomposition (SVD) to help determine filter parameters and to interpret the properties of the reconstructed volumes.
What problem does this paper attempt to address?