Initial evaluation of a direct 4D PET parametric image reconstruction method

jianhua yan,beata planetawilson,richard e carson
2009-01-01
Journal of Nuclear Medicine
Abstract:1462 Objectives The goal of this study is to evaluate an algorithm for direct reconstruction of PET parametric images from a one-tissue (1T) kinetic model. Methods Replicates of a 10-cm spherical phantom with 3 regions (Gray matter (GM), White matter(WM) and Basal Ganglia(BG)) were simulated; each comprised a 1-h list mode file (~4x108 events) based on a measured input function, typical kinetic parameters, and C-11 decay. Parametric images were produced with a frame-based method (FM) and the direct 4D method. The FM reconstructed individual frames (21 frames: 6x30s, 3x60s, 2x120s, 10x300s) and estimated each voxel9s kinetic parameters from its time-activity curve with weighted least squares (weights based on noise equivalent counts). The 4D method, an extension of the MOLAR algorithm, was derived from a new likelihood function which incorporated the 1T model into the physical projection model, and employed a novel EM algorithm to estimate the kinetic parameters directly from the list-mode data. Both methods used ordered subsets (30 subsets, 2 iterations). Evaluation was based on each region’s bias and coefficient of variation (COV) across replicates. Results Percent bias of both methods were small, with Conclusions This study indicates that the direct 4D reconstruction method is a promising approach to reduce noise in parametric images. Additional evaluation both by simulation and with real data is required.