A High Accuracy Technique to Measure the Electrical Conductivity of Liquids Using Small Test Samples

Jianping Wu,J. P. W. Stark
DOI: https://doi.org/10.1063/1.2709573
IF: 2.877
2007-01-01
Journal of Applied Physics
Abstract:A large quantity of a sample is usually required in electrical conductivity measurements of fluids. If precise electrical conductivity measurements are to be performed while using small test samples, it is necessary to study the effects of both fringe electric fields and interfacial properties. These effects were investigated here by using a novel technique termed the triangular waveform voltage method. The results using this technique revealed: (1) the fringe electric fields may be identified as an increase in apparent electrode area; and (2) the interfacial effects may be simply treated as an interfacial resistance. It was found that the increase of the electrode area due to the fringe electric fields may be reduced to below 1% at small electrode separations; the interfacial resistance is largely dependent of measurement frequency, electrode separation, and the electrical conductivity of the liquid. In general we show that the impact of interfacial resistance can be significantly reduced and the error in the electrical conductivity measurements for small liquid samples can be reduced from 40% to become less than ±1% in the tested NaCl(aq) solutions, spanning three decades of concentration, by adopting this new method.
What problem does this paper attempt to address?