Biomechanical Study of Unilateral Pedicle Screw Combined with Contralateral Translaminar Facet Screw in Transforaminal Lumbar Interbody Fusion.

Beier Luo,Ming Yan,Jinghui Huang,Wei Duan,Zhiping Huang,Jianting Chen,Zhuojing Luo,Qingan Zhu,Ming Li
DOI: https://doi.org/10.1016/j.clinbiomech.2015.05.009
IF: 2.034
2015-01-01
Clinical Biomechanics
Abstract:Background: The biomechanical stability of unilateral pedicle screw (UPS) combined with contralateral translaminar facet screw (TLFS), especially long-term stability, still needs to be compared to traditional UPS or bilateral pedide screws (BPSs) in details.Methods: Twenty-four porcine spines (L2-L5) were tested for flexibility with pure moments of 5 Nm under intact status and transforaminal lumbar interbody fusion status using UPS + TLFS, UPS or BPS at L3-L4. After short-term (3 cycles) and long-term cyclic loading (18,000 cycles), the range of motion was obtained and analyzed for single-level constructs in flexion/extension, lateral bending and axial rotation modes. In addition, the relative displacement of contralateral articular processes was recorded in a real time fashion.Findings: The range of motion was significantly reduced in all instrumented constructs. In all movement directions, UPS + TLFS achieved similar range of motion to BPS after short and long-term loading, which were significantly lower than that in UPS. A significantly larger displacement of contralateral articular process was recorded in UPS than UPS + TLFS and BPS during extension/flexion, lateral bending and axial rotation, suggesting its compromised stability.Interpretation: The hybrid construct of UPS + TLFS showed instant and long-term equivalent biomechanical ability to that of traditional BPS, making it an alternative option to BPS that could be less invasive while maintains a stable and effective instrumentation. (C) 2015 Elsevier Ltd. All rights reserved.
What problem does this paper attempt to address?