Photoionization mass spectrometer for studies of flame chemistry with a synchrotron light source

terrill a cool,andrew mcilroy,fei qi,phillip r westmoreland,lionel poisson,darcy s peterka,musahid ahmed
DOI: https://doi.org/10.1063/1.2010307
IF: 1.6
2005-01-01
Review of Scientific Instruments
Abstract:A flame-sampling molecular-beam photoionization mass spectrometer, recently designed and constructed for use with a synchrotron-radiation light source, provides significant improvements over previous molecular-beam mass spectrometers that have employed either electron-impact ionization or vacuum ultraviolet laser photoionization. These include superior signal-to-noise ratio, soft ionization, and photon energies easily and precisely tunable [E/Delta E(FWHM)approximate to 250-400] over the 7.8-17-eV range required for quantitative measurements of the concentrations and isomeric compositions of flame species. Mass resolution of the time-of-flight mass spectrometer is m/Delta m=400 and sensitivity reaches ppm levels. The design of the instrument and its advantages for studies of flame chemistry are discussed. (c) 2005 American Institute of Physics.
What problem does this paper attempt to address?