A Simulation and Experimental Study of Long Cortical Bones Fracture Evaluation Using Lamb Waves

Kailiang Xu,Runxin He,Dean Ta,Yixian Qin,Peng Sun,Weiqi Wang
DOI: https://doi.org/10.1121/1.4708845
2012-01-01
The Journal of the Acoustical Society of America
Abstract:Ultrasonic guided waves are promising to evaluate fractured long cortical bones. Guided modes conversion, occurred always on the sites of fractures, contains rich information of bone geometric and material properties. The study is to analyze the correlation between Lamb modes conversion and crack depth of fractured long bones. Axial experiments were designed to investigate the influences of different deep fractures in steel-made bone phantoms and sheep diaphyseal tibias over Lamb modes propagation, especially modes conversion. For comparison, the phantoms were modeled by two dimension finite-difference time-domain method. Group velocities and energies of S0, A0 and converted modes were extracted and analyzed under varying crack depth conditions. Being consistent with the simulation, experimental data showed obvious modes conversions occur between S0 and A0 modes. The fracture positions can be predicted from converted modes velocities. It illustrated the modes conversions become gradually observable with the fracture depth increasing, which can be indicated by the parameter, converted energy proportion (EP). The sensitivity and practicality of EP to assess bone fractures were validated in quantitative analysis of phantoms, simulations and sheep tibias experiments in vitro. In conclusion, ultrasonic guided wave is a reasonable method for long cortical bone fracture detection and evaluation.
What problem does this paper attempt to address?