Wearable scanning photoacoustic brain imaging in behaving rats.

Jianbo Tang,Xianjin Dai,Huabei Jiang
DOI: https://doi.org/10.1002/jbio.201500311
IF: 3.3899
2016-01-01
Journal of Biophotonics
Abstract:A wearable scanning photoacoustic imaging (wPAI) system is presented for noninvasive brain study in behaving rats. This miniaturized wPAI system consists of four pico linear servos and a single transducer-based PAI probe. It has a dimension of 50 mm x 35 mm x 40 mm, and a weight of 26 g excluding cablings. Phantom evaluation shows that wPAI achieves a lateral resolution of similar to 0.5 mm and an axial resolution of similar to 0.1 mm at a depth of up to 11 mm. Its imaging ability is also tested in a behaving rat, and the results indicate that wPAI is able to image blood vessels at a depth of up to 5 mm with intact scalp and skull. With its noninvasive, deep penetration, and functional imaging ability in behaving animals, wPAI can be used for behavior, cognition, and preclinical brain disease studies.
What problem does this paper attempt to address?