Computer-Simulated Biopsy Marking System for Endoscopic Surveillance of Gastric Lesions: A Pilot Study

Weiling Hu,Bin Wang,Leimin Sun,Shujie Chen,Liangjing Wang,Kan Wang,Jiaguo Wu,John J Kim,Jiquan Liu,Ning Dai,Huilong Duan,Jianmin Si
DOI: https://doi.org/10.1155/2015/197270
2015-01-01
BioMed Research International
Abstract:Endoscopic tattoo with India ink injection for surveillance of premalignant gastric lesions is technically cumbersome and may not be durable. The aim of the study is to evaluate the accuracy of a novel, computer-simulated biopsy marking system (CSBMS) developed for the endoscopic marking of gastric lesions. Twenty-five patients with history of gastric intestinal metaplasia received both CSBMS-guided marking and India ink injection in five points in the stomach at index endoscopy. A second endoscopy was performed at three months. Primary outcome was accuracy of CSBMS (distance between CSBMS probe-guided site and tattoo site measured by CSBMS). The mean accuracy of CSBMS at angularis was5.3±2.2 mm, antral lesser curvature5.7±1.4 mm, antral greater curvature6.1±1.1 mm, antral anterior wall6.9±1.6 mm, and antral posterior wall6.9±1.6 mm. CSBMS (2.3±0.9versus12.5±4.6seconds;P=0.02) required less procedure time compared to endoscopic tattooing. No adverse events were encountered. CSBMS accurately identified previously marked gastric sites by endoscopic tattooing within 1 cm on follow-up endoscopy.
What problem does this paper attempt to address?