Evaluation of Reekrossballoon catheter to treat long-segment occlusion of iliofemoral artery]

Yu Chen,Changwei Liu,Zhili Liu,Weiwei Wu,Rong Zeng,Xiaojun Song,Bao Liu
DOI: https://doi.org/10.3760/cma.j.issn.0376-2491.2015.01.013
2015-01-01
Abstract:The purpose of this study was to examine the safety and efficacy of the Reekross balloon catheter for recanalization of long CTO of iliofemoral arteries.A retrospective analysis of patients with long CTO of iliofemoral arteries who underwent angioplasty with Reekross balloon catheter between January 2011 and May 2013 was performed. All lesions were TASC II C and D. Sixty-eight patients(75 limbs) were enrolled with mean age of (69 ± 12) years, 53 male and 15 female. The limbs of Rutherford III, IV, V and VI were 41, 13, 19 and 2. Technical successrate, ankle brachial index (ABI), complication rate, and cumulative primary patency rate were evaluated.Technical success ratewas 90.7%. ABI increased from pre-procedure (0.41 ± 0.12) to 3-day post-procedure (0.77 ± 0.20) (t = -7.263, P < 0.01). Follow-up was achieved in 56 cases (61 limbs) from 12 to 40 months, with a mean of (26.3 ± 7.1) months. Primary patency rates at 3 months, 6 months, 12 months and 24 months were (98.4 ± 1.6) %, (86.7 ± 4.4)%, (80.0 ± 5.2) % and ( 69.4 ± 6.3) %.The Reekross balloon catheter is safe and effective for successful recanalization of CTO of iliofemoral artery.
What problem does this paper attempt to address?