Using Size-Selected Gold Clusters on Graphene Oxide Films to Aid Cryo-Transmission Electron Tomography Alignment

Kenton P. Arkill,Judith M. Mantell,Simon R. Plant,Paul Verkade,Richard E. Palmer
DOI: https://doi.org/10.1038/srep09234
IF: 4.6
2015-01-01
Scientific Reports
Abstract:A three-dimensional reconstruction of a nano-scale aqueous object can be achieved by taking a series of transmission electron micrographs tilted at different angles in vitreous ice: cryo-Transmission Electron Tomography. Presented here is a novel method of fine alignment for the tilt series. Size-selected gold clusters of ~2.7 nm (Au 561 ± 14 ), ~3.2 nm (Au 923 ± 22 ) and ~4.3 nm (Au 2057 ± 45 ) in diameter were deposited onto separate graphene oxide films overlaying holes on amorphous carbon grids. After plunge freezing and subsequent transfer to cryo-Transmission Electron Tomography, the resulting tomograms have excellent (de-)focus and alignment properties during automatic acquisition. Fine alignment is accurate when the evenly distributed 3.2 nm gold particles are used as fiducial markers, demonstrated with a reconstruction of a tobacco mosaic virus. Using a graphene oxide film means the fiducial markers are not interfering with the ice bound sample and that automated collection is consistent. The use of pre-deposited size-selected clusters means there is no aggregation and a user defined concentration. The size-selected clusters are mono-dispersed and can be produced in a wide size range including 2–5 nm in diameter. The use of size-selected clusters on a graphene oxide films represents a significant technical advance for 3D cryo-electron microscopy.