Dry-contact Microelectrode Membranes for Wireless Detection of Electrical Phenotypes in Neonatal Mouse Hearts

Yu Zhao,Hung Cao,Tyler Beebe,Hemin Zhang,Xiaoxiao Zhang,Honglong Chang,Oscar Scremin,Ching-Ling Lien,Yu-Chong Tai,Tzung K. Hsiai
DOI: https://doi.org/10.1007/s10544-014-9912-y
2015-01-01
Biomedical Microdevices
Abstract:Continuous monitoring of aberrant electrical rhythms during heart injury and repair requires prolonged data acquisition. We hereby developed a wearable microelectrode membrane that could be adherent to the chest of neonatal mice for in situ wireless recording of electrocardiogram (ECG) signals. The novel dry-contact membrane with a meshed parylene-C pad adjacent to the microelectrodes and the expandable meandrous strips allowed for varying size of the neonates. The performance was evaluated at the system level; specifically, the ECG signals (μV) acquired from the microelectrodes underwent two-stage amplification, band-pass filtering, and optical data transmission by an infrared Light Emitting Diode (LED) to the data-receiving unit. The circuitry was prototyped on a printed circuit board (PCB), consuming less than 300 μW, and was completely powered by an inductive coupling link. Distinct P waves, QRS complexes, and T waves of ECG signals were demonstrated from the non-pharmacologically sedated neonates at ~600 beats per minutes. Thus, we demonstrate the feasibility of both real-time and wireless monitoring cardiac rhythms in a neonatal mouse (17–20 mm and <1 g) via dry-contact microelectrode membrane; thus, providing a basis for diagnosing aberrant electrical conduction in animal models of cardiac injury and repair.
What problem does this paper attempt to address?