A carboxylated graphene and aptamer nanocomposite-based aptasensor for sensitive and specific detection of hemin.

Ling Gao,Yinghong Xiao,Yanping Wang,Xue Chen,Bo Zhou,Xiaodi Yang
DOI: https://doi.org/10.1016/j.talanta.2014.09.010
IF: 6.1
2015-01-01
Talanta
Abstract:A unique nanocomposite was crafted by grafting hemin-binding-aptamer (HBA) onto carboxylated graphene (COO-GR). Infrared spectroscopy, Raman spectroscopy and diffuse reflectance spectra suggested that –NHCO– covalent bonds were formed between HBA and COO-GR. The resulting COO-GR/HBA functionalized electrode was used as a novel label-free biosensor. The square wave voltammetry was employed to realize the selective and specific detection of hemin. The obtained aptasensor possessed excellent performance with a detection limit of 0.64nmolL−1 (S/N=3) and a linear range from 1 to 150nmolL−1. Moreover, COO-GR was shown to be a promising candidate in making aptasensors, carrying advantages over graphene in terms of the simplicity of sensor preparation and the reduction of background noise.