A disposable electrochemical sensor for simultaneous determination of norepinephrine and serotonin in rat cerebrospinal fluid based on MWNTs-ZnO/chitosan composites modified screen-printed electrode.

Yuting Wang,Shi Wang,Li Tao,Qing Min,Jin Xiang,Qiman Wang,Jianmei Xie,Yuan Yue,Shuangchan Wu,Xiaofei Li,Hong Ding
DOI: https://doi.org/10.1016/j.bios.2014.09.099
IF: 12.545
2015-01-01
Biosensors and Bioelectronics
Abstract:A new electrochemical sensor for simultaneous determination of norepinephrine (NE) and serotonin (5-HT) was fabricated. The electrochemical behavior of NE and 5-HT were investigated using CV and SWV at the MWNTs-ZnO/chitosan composites modified screen-printed electrode (MWNTs-ZnO/chitosan SPE). The results showed that the current responses of NE and 5-HT greatly enhanced due to the high catalytic activity of composites. The peak potentials of NE and 5-HT were separated at about 90mV and 280mV, respectively. The peak currents of NE and 5-HT were linearly dependent on their concentrations in the range of 0.5–30μM and 0.05–1μM, with the limit of detection of 0.2μM and 0.01μM, respectively. The modified electrode can be stored stably for at least 3 month at 4°C in a refrigerator. Furthermore, the modified electrode was successfully applied to detect the level of NE and 5-HT in rat cerebrospinal fluid (CSF) with excellent selectivity and sensitivity.