Determination of Autoinducer-2 in Biological Samples by High-Performance Liquid Chromatography with Fluorescence Detection Using Pre-Column Derivatization.

Xiang-Ning Song,Hai-Bin Qiu,Xiang Xiao,Yuan-Yuan Cheng,Wen-Wei Li,Guo-Ping Sheng,Xiao-Yan Li,Han-Qing Yu
DOI: https://doi.org/10.1016/j.chroma.2014.07.103
2014-01-01
Abstract:Autoinducer-2 (AI-2), as a small-molecular-weight organic molecule secreted and perceived by various bacteria, enables intra- and inter-species communications. Quantitative determination of AI-2 is essential for exploring the bacterial AI-2-related physiological and biochemical processes. However, current strategies for sensitive detection of AI-2 require sophisticated instruments and complicated procedures. In this work, on the basis of the derivatization of AI-2 with 2,3-diaminonaphthalene, a simple, sensitive and cost-effective high-performance liquid chromatography with fluorescence detector (HPLC-FLD) method is developed for the quantitative detection of AI-2. Under the optimized conditions, this method had a broad linear range of 10-14,000 ng/ml(R-2=0.9999), and a low detection limit of 1.0 ng/ml. Furthermore, the effectiveness of this approach was further validated through measuring the AI-2 concentrations in the cell-free culture supernatants of both Escherichia coli and Vibrio harveyi. (C) 2014 Elsevier B.V. All rights reserved.
What problem does this paper attempt to address?