Interfacing Microsampling Droplets and Mass Spectrometry by Paper Spray Ionization for Online Chemical Monitoring of Cell Culture

Wu Liu,Niejun Wang,Xuexia Lin,Yuan Ma,Jin-Ming Lin
DOI: https://doi.org/10.1021/ac501678q
IF: 7.4
2014-01-01
Analytical Chemistry
Abstract:In this work, the establishment of a microdialysis-paper spray ionization-mass spectrometry (MS) system was described. A homemade microdialysis module was employed for sampling, and microdroplets were generated at the outlet of the capillary conducting the dialysate. Online MS analysis of each microdroplet was immediately accomplished, interfacing by paper spray ionization. Analytical performance of the method was investigated and improved through the introduction of thinner capillary tubes and the optimization of spray solvent and paper substrate. For microdroplets with concentrated salt at 50 nL, the limit of detection at 0.8 ppm (or 40 pg absolute) and a highest resolution at about 1.5 s were achieved. The integrated system was applied into the online monitoring of glucose concentration in cell culture mediums. A satisfactory linearity of the calibration curve between the relative MS intensity and the glucose concentration was observed. Furthermore, as a model, hormone regulation of the glucose concentration was investigated. This work demonstrated the potential application of the label-free, online "MS sensor" into studies on cellular metabolism.
What problem does this paper attempt to address?