An ultra-high vacuum electrochemical flow cell for in situ/operando soft X-ray spectroscopy study.

Debajeet K Bora,Per-Anders Glans,John Pepper,Yi-Sheng Liu,Chun Du,Dunwei Wang,J-H Guo
DOI: https://doi.org/10.1063/1.4870795
IF: 1.6
2014-01-01
Review of Scientific Instruments
Abstract:An in situ flow electrochemical cell has been designed and fabricated to allow better seal under UHV chamber thus to achieve a good signal to noise ratio in fluorescence yield detection of X-ray absorption spectra for spectroelectrochemical study. The cell also stabilizes the thin silicon nitride membrane window in an effective manner so that the liquid cell remains intact during X-ray absorption experiments. With the improved design of the liquid cell, electrochemical experiments such as cyclic voltammetry have been performed for 10 cycles with a good stability of sample window. Also an operando electrochemical experiment during photoelectrochemistry has been performed on n-type hematite electrode deposited on silicon nitride window. The experiment allows us to observe the formation of two extra electronic transitions before pre edge of O K-edge spectra. (C) 2014 AIP Publishing LLC.
What problem does this paper attempt to address?