Microfluidic-SERS Devices for One Shot Limit-of-detection.

Donghyuk Kim,Antonio R. Campos,Ashish Datt,Zhe Gao,Matthew Rycenga,Nathan D. Burrows,Nathan G. Greeneltch,Chad A. Mirkin,Catherine J. Murphy,Richard P. Van Duyne,Christy L. Haynes
DOI: https://doi.org/10.1039/c4an00357h
2014-01-01
Abstract:Microfluidic sensing platforms facilitate parallel, low sample volume detection using various optical signal transduction mechanisms. Herein, we introduce a simple mixing microfluidic device, enabling serial dilution of introduced analyte solution that terminates in five discrete sensing elements. We demonstrate the utility of this device with on-chip fluorescence and surface-enhanced Raman scattering (SERS) detection of analytes, and we demonstrate device use both when combined with a traditional inflexible SERS substrate and with SERS-active nanoparticles that are directly incorporated into microfluidic channels to create a flexible SERS platform. The results indicate, with varying sensitivities, that either flexible or inflexible devices can be easily used to create a calibration curve and perform a limit of detection study with a single experiment.
What problem does this paper attempt to address?