A Rapid and High‐precision Method for Sulfur Isotope Δ34s Determination with a Multiple‐collector Inductively Coupled Plasma Mass Spectrometer: Matrix Effect Correction and Applications for Water Samples Without Chemical Purification

An-Jun Lin,Tao Yang,Shao-Yong Jiang
DOI: https://doi.org/10.1002/rcm.6838
IF: 2.586
2014-01-01
Rapid Communications in Mass Spectrometry
Abstract:RATIONALEPrevious studies have indicated that prior chemical purification of samples, although complex and time‐consuming, is essential in obtaining precise and accurate results for sulfur isotope ratios using multiple‐collector inductively coupled plasma mass spectrometry (MC‐ICP‐MS). In this study, we introduce a new, rapid and precise MC‐ICP‐MS method for sulfur isotope determination from water samples without chemical purification.METHODSThe analytical work was performed on an MC‐ICP‐MS instrument with medium mass resolution (m/Δm ~ 3000). Standard‐sample bracketing (SSB) was used to correct samples throughout the analytical sessions. Reference materials included an Alfa‐S (ammonium sulfate) standard solution, ammonium sulfate provided by the lab of the authors and fresh seawater from the South China Sea. A range of matrix‐matched Alfa‐S standard solutions and ammonium sulfate solutions was used to investigate the matrix (salinity) effect (matrix was added in the form of NaCl). A seawater sample was used to confirm the reliability of the method.RESULTSUsing matrix‐matched (salinity‐matched) Alfa‐S as the working standard, the measured δ34S value of AS (−6.73 ± 0.09‰) was consistent with the reference value (−6.78 ± 0.07‰) within the uncertainty, suggesting that this method could be recommended for the measurement of water samples without prior chemical purification. The δ34S value determination for the unpurified seawater also yielded excellent results (21.03 ± 0.18‰) that are consistent with the reference value (20.99‰), thus confirming the feasibility of the technique.CONCLUSIONSThe data and the results indicate that it is feasible to use MC‐ICP‐MS and matrix‐matched working standards to measure the sulfur isotopic compositions of water samples directly without chemical purification. In comparison with the existing MC‐ICP‐MS techniques, the new method is better for directly measuring δ34S values in water samples with complex matrices; therefore, it can significantly accelerate analytical turnover. Copyright © 2014 John Wiley & Sons, Ltd.
What problem does this paper attempt to address?