Modelling and Control of a Five‐degrees‐of‐freedom Pneumatically Actuated Magnetic Resonance‐compatible Robot

Shan Jiang,Wenhao Feng,Jinlong Lou,Zhiyong Yang,Jun Liu,Jun Yang
DOI: https://doi.org/10.1002/rcs.1524
2013-01-01
International Journal of Medical Robotics and Computer Assisted Surgery
Abstract:BackgroundMagnetic resonance imaging (MRI) combined with robotic assistance has the potential to improve on clinical outcomes of biopsy and local treatment of prostate cancer. This paper introduces a pneumatically actuated surgical robotic system for prostatic interventions under MRI guidance. MethodsA mechanical dynamics model was established using the Lagrange method and simulated using MATLAB software. A pneumatic actuator model was built through the use of system identification, and stability analysis was proposed based on the Nyquist criterion. In addition, a proportional-integral-derivative controller was implemented on the robot for accurate position control. ResultsSimulation results showed that the steady-state error of a step response was 0.3mm and the tracking error of a 1.0Hz sinusoidal response was 1.5mm. Experimental results showed that the insertion precision was 0.72mm in the general laboratory environment. ConclusionsThe results demonstrated that the robotic system with the designed controller was effective in both position control and trajectory tracking. Copyright (c) 2013 John Wiley & Sons, Ltd.
What problem does this paper attempt to address?