Application Evaluation of the Maxmat PL Multi-Function Blood Coagulation Analyzer
Han XU,Cheng-bin WANG,Ping DI,Jian LI
2014-01-01
Abstract:Objective To evaluate the performance of the Maxmat PL multi-function blood coagulation analyzer .Methods The assay precision ,linearity,carry-over,comparability and correlation were assessed on the Maxmat PL multi-function blood coagulation analyzer according to national industry standards .Results Intra-assay precision of PT,INR,aPTT,Fbg,TT,D-dimer of normal quality control CV values were 1.66, 1.68,1.81,1.52,1.21,11.55,respectively;The abnormal quality control CV values were 1.11,1.21,1.46, 3.58,0.35,6.14,respectively.Inter-assay precision of PT,INR,aPTT,Fbg,TT,D-dimer of normal quality control CV values were 1.51,2.01,1.93,0.99,1.24,16.03,respectively;The abnormal quality control CV values were 1.03,1.11,1.26,3.16,0.22,5.03,respectively.The results of PT,INR,aPTT,Fbg,TT were closely related to that measured on the french STA-R coagulation analyzer ( r were 0.996,0.996,0.958, 0.974,0.982,respectively).Compared with the French STA-R coagulation analyzer,D-dimer positive agree-ment was 96.8%,the correlative of the PT,aPPT,Fbg were 0.973,0.944,0.962,respectively.Conclusion The Maxmat PL multi-function blood coagulation analyzer verified to be a reliable instrument with convenient performance characteristics .The analyzer improves and simplifies the laboratory workload .