Automated Online Pretreatment and Cleanup Recycle Coupled with High-Performance Liquid Chromatography-Mass Spectrometry for Determination of Deca-Bromodiphenyl Ether in Human Serum.

Xuexia Lin,Hai-Fang Li,Xiangwei He,Yuki Hashi,Jin-Ming Lin,Zhihua Wang
DOI: https://doi.org/10.1002/jssc.201200211
IF: 3.614
2012-01-01
Journal of Separation Science
Abstract:Automated online SPE-HPLC-MS was established for the determination of deca-bromodiphenyl ether in human serum. The online SPE with large volume injection was utilized to enhance the sensitivity. Online SPE with dilution line greatly decreased matrices effect, which enabled serum samples to be injected directly into pre-column. Washing line was designed for the system to solve the serious residual phenomenon and reduce the risk of sample wastage and contamination. Under the optimized conditions, the linear of the method was in the range 0.1-10 ng/mL with the LOD of 0.026 ng/mL. The recoveries of serum samples spiked with deca-bromodiphenyl ether at 0.5 ng/mL was in the range from 83.30 to 102.7% with RSD in interday less than 8.67%. The satisfactory results demonstrated that the method of online sample pretreatment and cleanup recycle were reliable for human serum analysis.