A Negative-pressure Flow-injection Micro-electrode System for Rapid and Simultaneous Determination of Four Electrolytes in Human Serums

Li Yong-Sheng,Zhou Lei,Zhang Hui-Sheng,Lai Zhen-Yao,Gao Xiu-Feng
DOI: https://doi.org/10.2116/analsci.28.781
IF: 1.967
2012-01-01
Analytical Sciences
Abstract:An automatic system for simultaneous determination of K + , Na + , Cl − and Ca 2+ in serums was established on the basis of a negative-pressure flow-injection using micro-electrodes, and parameters of the system were optimized. The total ionic strength adjustment buffer consisted of 0.25 mmol L −1 K + , 48.6 mmol L −1 Na + , 2.56 mmol L −1 Cl − , 0.25 mmol L −1 Ca 2+ and 23 mmol L −1 Na 2 B 4 O 7 -H 3 BO 3 ; the flow rate was 1.58 mL min −1 , the sampling volume was 45 μL, and the mixing coil length was 30 cm. The system could conduct 480 detections h −1 , and its RSD was less than 1.6%. Recoveries were 97.3 – 103.6%, and linear responses were 2.0 – 22.0 mmol L −1 for K + , 89.6 – 253.0 mmol L −1 for Na + , 20.1 – 248.2 mmol L −1 for Cl − , and 0.35 – 10.0 mmol L −1 for Ca 2+ , respectively. The ion concentration ranges were in ranges of human serum electrolytes. The system features minimized sorption of fibrin in sensors and prolonged sensor life.