[Effect of Bicortical Orthodontic Microimplant on Tooth Movement in Dogs].

Jian-chao Wu,Ji-na Huang,Shi-fang Zhao
DOI: https://doi.org/10.3785/j.issn.1008-9292.2006.05.005
2006-01-01
Abstract:OBJECTIVE:To evaluate the potential anchorage of bicortical microimplant for tooth movement.METHODS:Five bicortical microimplants were inserted in the interradicular area of the second premolar (P2) in one side of the mandible (test side), 5 monocortical microimplant in the contralateral region (control side) in 5 beagle dogs. A total of 100 g force was generated between the implant and the fourth premolar (P4) in two sides. M1-P4, CA-IB and CA-M1 were measured biweekly. At the end of loading, specimens with P4 segments were cut and grounded to 70 microm to calculate the bone-to-implant contact (BIC).RESULT:All the bicortical microimplants remained stable during the treatment periods, while 1 monocortical microimplant was lost within 1 week. Mesial movement of P4 was ( 3.92+/-0.22) mm in the test side, (2.03+/-0.15) mm in the control side (P<0.05). Analysis showed no difference of the BIC between the bicortical microimplants and the monocortical microimplants.CONCLUSION:The bicortical microimplant may be used as orthodontic anchorage for mesial movement of posterior tooth.