Simultaneous Determination of Bisphenol A and Tetrabromobisphenol A in Plastic Products Using High Performance Liquid Chromatography-Mass Spectrometry

Huang Shaochan,Hang Yiping
DOI: https://doi.org/10.3724/sp.j.1123.2010.00863
2010-01-01
Chinese Journal of Chromatography
Abstract:A method for the determination of bisphenol A (BPA) and tetrabromobisphenol A (TBBP-A) in plastic products using high performance liquid chromatography-mass spectrometry (HPLC-MS) was developed with the ultrasonic extraction. The conditions of sample pretreatment, chromatography and mass-spectrometry parameters were studied systematically. The research showed that the extraction of the BPA and TBBP-A can reach a good performance with 20 mL dichloromethane at 50 degrees C under ultrasonic extraction for 60 min. The BPA and TBBP-A were separated and detected using HPLC-MS with the mobile phase of methanol and water. The linearity range of the method was 0.1 - 2.0 mg/L. The detection limits were 0.01 mg/kg for BPA and 0.02 mg/kg for TBBP-A. The recoveries for the spiked samples were 85.4%-97.6%. This method is time-saving; easy-handling; practicable and sensitive; and it can be applied to determine the residues of BPA and TBBP-A in plastic products.
What problem does this paper attempt to address?