Label-free Biosensing Using a Photonic Crystal Structure in a Total-Internal-reflection Geometry

Yunbo Guo,Jing Yong Ye,Charles Divin,Thommey P. Thomas,Andrzej Myc,Tommaso F. Bersano-Begey,James R. Baker,Theodore B. Norris
DOI: https://doi.org/10.1117/12.808369
2009-01-01
Abstract:A novel optical biosensor using a one-dimensional photonic crystal structure in a total-internal-reflection geometry (PC-TIR) is presented and investigated for label-free biosensing applications. This simple configuration forms a micro Fabry-Perot resonator in the top layer which provides a narrow optical resonance to enable label-free, highly sensitive measurements for the presence of analytes on the sensing surface or the refractive index change of the surrounding medium in the enhanced evanescent field; and at the same time it employs an open sensing surface for real-time biomolecular binding detection. The high sensitivity of the sensor was experimentally demonstrated by bulk solvent refractive index changes, ultrathin molecular films adsorbed on the sensing surface, and real-time analytes binding, measuring both the spectral shift of the photonic crystal resonance and the change of the intensity ratio in a differential reflectance measurement. Detection limits of 7x10(-8) RIU for bulk solvent refractive index, 6x10(-5) nm for molecular layer thickness and 24 fg/mm(2) for mass density were obtained, which represent a significant improvement relative to state-of-the-art surface-plasmon-resonance (SPR)-based systems. The PC-TIR sensor is thus seen to be a promising technology platform for high sensitivity and accurate biomolecular detection.
What problem does this paper attempt to address?