A New Method for the Analysis of Beta(2)-Agonists in Human Urine by Pressure-Assisted Capillary Electrochromatography Coupled with Electrospray Ionization-Mass Spectrometry Using a Silica-Based Monolithic Column.

Minghua Lu,Lan Zhang,Xin Li,Qiaomei Lu,Guonan Chen,Zongwei Cai
DOI: https://doi.org/10.1016/j.talanta.2010.03.020
IF: 6.1
2010-01-01
Talanta
Abstract:A new pressure-assisted capillary electrochromatography coupled with electrospray ionization-mass spectrometry method using a silica-based monolithic column as separation media was developed for the analysis of β2-agonists in human urine. Experimental conditions including the mobile phase, separation voltage, assisted pressure, and sheath liquid were optimized for the analysis: mobile phase composed of 82% (v/v) ACN and 18% (v/v) 20mmol/L ammonium acetate (pH 6.0); separation voltage 25kV; assisted pressure 2bar; and the sheath liquid consisting of 7.5mmol/L acetic acid in isopropanol/water 50/50% (v/v) that was delivered at a flow rate of 3.0μL/min. Six β2-agonists were separated within 12.5min with LODs (defined as S/N=3) in the range of 0.25–2.0ng/mL. The absolute LODs of the developed method for analyzing six β2-agonists ranged from 5.75 to 46.0fg. Method repeatability of run-to-run and column-to-column was satisfactory. The recovery obtained from the analysis of spiked urine samples was between 88.2% and 106% with RSDs lower than 6.68%. The method was successfully applied to the analysis of real urine sample from volunteers.