Abstract:Our method is composed of three steps: The 3D image is projected along an initial direction, and the needle is segmented in the projected image. Using the projection direction and detected 2D needle direction, a plane containing the needle, called the needle plane, is determined. Secondly, the 3D image is projected in the direction perpendicular to the normal of the needle plane and Step 1 is repeated. If the needle direction in the projected image is horizontal, the needle plane is correct, otherwise Steps 1 & 2 are repeated until a correct needle plane has been found. Thirdly, the 3D image is projected along the normal direction of the needle plane. In this projected image, we determine the endpoints of the needle in the projection coordinates. Then, using the relationship between the projection coordinates and the 3D image coordinates, we determine the end points of the needle in the 3D US image. Experiments with agar and turkey phantom 3D US images demonstrated that our method can segment the needle from 3D US images accurately and robustly with a speed of 13 fps on a 1.8GHz PC computer.
What problem does this paper attempt to address?