Electrical Conductivity Measurement in Thiel-Embalmed Tissue Model: Relevance to Radiofrequency Ablation

Zhigang Wang,Hongyan Luo,M. Nick,M. Gueorguieva,P. Andre,R. T. Baker,D. McLean,S. Brown,A. Cuschieri
DOI: https://doi.org/10.1049/el.2014.1934
2014-01-01
Electronics Letters
Abstract:Thiel-embalmed human cadaveric specimens are used widely for biomedical scientific investigation. This Letter reports electrical impedance measurements of Thiel-embalming solutions from 1 Hz to 32 MHz. Compared with other solutions studied, Thiel solution has the lowest impedance throughout the test spectrum. The electrical conductivity of a Thiel-embalmed liver sample exhibited a relatively flat frequency response from 100 to 500 kHz, which is the frequency range used for electro-surgery and radiofrequency tumour ablation (RFA). It measured 5 Sm−1 compared with 0.22 Sm−1 obtained from ex-vivo fresh pig liver. Using finite element modelling and experimental evaluation, the ablation zone obtained from the Thiel-embalmed liver sample was extremely small due to its much higher conductivity. Hence, it is concluded that Thiel-embalmed tissue cannot be used as a reliable model for RFA evaluation.
What problem does this paper attempt to address?