Minimal-scan Reconstruction Algorithms for Fan-Beam Diffraction Tomography and Their Analogy to Halfscan Fan-Beam CT

Xiaochuan Pan,Anastasio, M.A.
DOI: https://doi.org/10.1109/nssmic.2000.950035
2000-01-01
Abstract:In fan-beam diffraction tomography (DT), a scattering object is interrogated using a cylindrical acoustical or electromagnetic wavefield, and the scattered wavefield around the object is measured and used to reconstruct the refractive index distribution of the scattering object. Recently, there has been a strong interest in developing ultrasound-, microwave- and photon density-based DT systems as medical imaging modalities. In conventional fan-beam computed tomography (CT), it is possible to reduce the scanning time by use of the so-called minimal-scan approach (also referred to as the halfscan approach) in which one acquires the minimal-scan fan-beam sinogram at projection angles from zero to π plus the fan angle. It is widely believed that in fan-beam DT, measurements from a full angular range of 2 π around the scattering object are generally required to exactly reconstruct a complex-valued refractive index distribution. In this work the authors reveal that to perform an exact reconstruction, one needs measurements only over the angular range O⩽φ⩽φ min, where π<φmin⩽3π/2 is a specified function describing the fan-beam DT geometry. Based on this observation, the authors develop minimal-scan reconstruction algorithms for fan-beam DT. They also present a discussion of the conceptual similarities between the minimal-scan fan-beam DT and CT reconstruction problems