Dual-slope Capacitance to Digital Converter Integrated in an Implantable Pressure Sensing System

Sechang Oh,Yoonmyung Lee,Jingcheng Wang,Zhiyoong Foo,Yejoong Kim,David Blaauw,Dennis Sylvester
DOI: https://doi.org/10.1109/jssc.2015.2435736
IF: 5.4
2015-01-01
IEEE Journal of Solid-State Circuits
Abstract:This work presents a dual-slope capacitance to digital converter for pressure sensing. The design uses base capacitance subtraction with a configurable capacitor bank and dual precision comparators to improve energy efficiency, consuming 110nW with 9.7b ENOB and 0.85pJ/conv·step FoM. The converter is integrated with a pressure transducer, battery, processor, and radio to form a complete 1.4mm×2.8mm×1.6mm sensor system aimed at implantable devices. The system operates from a 3.6V battery.
What problem does this paper attempt to address?