Modeling Of Temperature Control For An Intraperitoneal Hyperthermic Chemotherapy Machine

Teng Wang,Xinjun Sheng,Jianwei Liu,Qiumeng Yang,Xiangyang Zhu
DOI: https://doi.org/10.1007/978-3-319-13963-0_34
2014-01-01
Abstract:Intraperitoneal hyperthermic chemotherapy (IHCT) combines the efficacies of hyperthermia and chemotherapy on tumor treatment, which achieves satisfactory curative effects in treating abdominal cancer and decreasing the recurrence rate after operation. To design an IHCT machine according to treatment requirements, the temperature control system of chemotherapy liquid is analyzed in detail in this study. A mathematical model on liquid heating process is obtained using the system identification technique. The structure of this model is simple and the fitting accuracy surpasses 92%. The efforts in this study can provide a theoretical basis for liquid temperature control of IHCT.
What problem does this paper attempt to address?