Passive Breath Gating Equipment for Cone Beam CT-guided RapidArc Gastric Cancer Treatments.

Weigang Hu,Guichao Li,Jinsong Ye,Jiazhou Wang,Jiayuan Peng,Min Gong,Xiaoli Yu,Matthew T. Studentski,Ying Xiao,Zhen Zhang
DOI: https://doi.org/10.1016/j.radonc.2014.11.046
IF: 6.901
2015-01-01
Radiotherapy and Oncology
Abstract:Background and purposeTo report preliminary results of passive breath gating (PBG) equipment for cone-beam CT image-guided gated RapidArc gastric cancer treatments.Material and methodsHome-developed PBG equipment integrated with the real-time position management system (RPM) for passive patient breath hold was used in CT simulation, online partial breath hold (PBH) CBCT acquisition, and breath-hold gating (BHG) RapidArc delivery. The treatment was discontinuously delivered with beam on during BH and beam off for free breathing (FB). Pretreatment verification PBH CBCT was obtained with the PBG-RPM system. Additionally, the reproducibility of the gating accuracy was evaluated.ResultsA total of 375 fractions of breath-hold gating RapidArc treatments were successfully delivered and 233 PBH CBCTs were available for analysis. The PBH CBCT images were acquired with 2–3 breath holds and 1–2 FB breaks. The imaging time was the same for PBH CBCT and conventional FB CBCT (60s). Compared to FB CBCT, the motion artifacts seen in PBH CBCT images were remarkably reduced. The average BHG RapidArc delivery time was 103s for one 270-degree arc and 269s for two full arcs.ConclusionsThe PBG-RPM based PBH CBCT verification and BHG RapidArc delivery was successfully implemented clinically. The BHG RapidArc treatment was accomplished using a conventional RapidArc machine with high delivery efficiency.
What problem does this paper attempt to address?