Femtosecond Broadband Fluorescence Upconversion Spectroscopy: Improved Setup and Photometric Correction.

X. -X. Zhang,C. Wuerth,L. Zhao,U. Resch-Genger,N. P. Ernsting,M. Sajadi
DOI: https://doi.org/10.1063/1.3597674
2011-01-01
Abstract:A setup for fluorescence upconversion spectroscopy (FLUPS) is described which has 80 fs temporal response (fwhm) for emission in the spectral range 425-750 nm. Broadband phase matching is achieved with tilted gate pulses at 1340 nm. Background from harmonics of the gate pulse is removed and sensitivity increased compared to previous designs. Photometric calibration of the upconversion process is performed with a set of fluorescent dyes. For Coumarin 153 in methanol the peak position, bandwidth, and asymmetry depending on delay time are reported.