Non-Invasive Measurement of Human Blood Protein Concentration Based on Dynamic Spectrum Method

LI Gang,MEN Jian-long,SUN Zhao-min,WANG Hui-quan,LIN Ling,ZHANG Bao-ju,WU Xiao-rong
DOI: https://doi.org/10.3969/j.issn.0493-2137.2011.01.017
2011-01-01
Abstract:Non-invasive determination of blood components concentration,as one of the medical analysis technique,has attracted much interest in recent years,but it has not yet been applied to the clinical field due to the effect of individual discrepancy and complicated measurement condition.To solve these problems,a non-invasive(NI)in vivo measurement technique of human blood proteins concentration using near-infrared dynamic spec-trum(DS)method was put forward.In vivo measurements were carried out on 110 volunteers,and biochemical ex-aminations of blood components contents,including hemoglobin(HGB),total protein(TP)and albumin(ALB)were conducted simultaneously to establish partial least squares(PLS) calibration models.Correlation coefficients between calibration sets and prediction sets were 0.999 4,0.922 1 and 0.956 0 respectively,and the average relative errors of prediction sets were 3.77%,3.49% and 3.40% respectively.The results show that the dynamic spectrum analysis can achieve high accuracy and has provided a promising NIR approach to the non-invasive measurement of human blood components.