Establishment and evaluation of time-resolved fluoroimmunoassay for quantitative detection of tissue factor pathway inhibitor-2

Fang-lin JIANG,Jing LIU,Hui-jun WANG,Jin ZHANG,Xin-ping LUO,Duan MA
DOI: https://doi.org/10.3969/j.issn.1672-8467.2011.01.013
2011-01-01
Abstract:Objective To establish a new and high sensitive method of time-resolved fluorescence immunoassay (TRFIA) for the quantification of human tissue factor pathway inhibitor-2 (TFPI-2). Methods Polyclonal antihuman TFPI-2 rabbit IgG was used as the capture antibody, monoclonal antibody 3C8 as the first antibody, hTFPI-2 expressed by E. coli as the standards, and Eu3+ -labelled goat anti-mouse IgG as the second antibody. Results The linear range of standard curve of TFPI-2 TRFIA was 0.1-50 ng/mL, and the sensitivity was 0.049 ng/mL. The intra- and inter- coefficients of variation were both <8.39%. The concentration of hTFPI-2 in plasma of 56 patients with coronary atherosclerotic heart disease (CHD) and 24 health controls were (10.58 ± 7.12) and (8.53 ± 2.32) ng/mL, respectively. Conclusions The TFPI-2 TRFIA kit we developed is more sensitive than ELISA kit, and it can be used to detect TFPI-2 in circulation.