Nasotracheal Intubation Using the Blind Intubation Device in Anaesthetised Adults with Mallampati Class 3

Yu Sun,Jin-Xing Liu,Ye-Sen Zhu,Hui Xu,Yan Huang,Hong Jiang
DOI: https://doi.org/10.1097/eja.0b013e328349a9f9
2011-01-01
European Journal of Anaesthesiology
Abstract:Context We hypothesised that the Blind Intubation Device (BID) would be effective for nasotracheal intubation (NTI) in anaesthetised adults with Mallampati class 3. We also hypothesised that BID may cause less haemodynamic changes due to the avoidance of direct stimulation induced by the Macintosh blade.Objective The purpose of the study was to compare the effectiveness of the BID with the Macintosh laryngoscope for NTI in anaesthetised adults with Mallampati class 3.Design A prospective randomised controlled study.Setting Operation unit in a University Hospital in Shanghai. Period of the study was from September to November 2010.Patients or other participants Mallampati class 3 adults requiring NTI for elective oral and maxillofacial surgery were randomly assigned to a BID group (n = 25) or a Macintosh laryngoscope group (ML group) (n = 25).Intervention After anaesthesia induction, patients were intubated by a single anaesthesiologist experienced in using both devices.Main outcome measures The mean arterial pressure (MAP) and heart rate (HR) were recorded at specific time points. NTI duration and success rate was compared. Epistaxis-associated and NTI-associated postoperative complications were assessed.Results Compared with baseline values, there was a significant increase in MAP in both the BID and ML groups which persisted significantly longer in the ML group. The BID group showed a significantly attenuated MAP value within 30-60 s. The difference between the maximum MAP and the post-induction value was significantly greater in the ML group than in the BID group (64.4 +/- 16.1 vs. 45.9 +/- 16.1 mmHg, P = 0.0003). Compared with baseline values, there was a significant increase in HR in both groups which persisted longer in the ML group. There was a significantly higher first attempt success rate in the BID group compared with the ML group (100 vs. 76%, respectively, P = 0.022). The NTI duration was 36 s [interquartile range (IQR) 32-40] in the BID group and 33 s (IQR 25.5-41.5) in the ML group. Epistaxis during NTI was less frequent and less severe in the BID group (P = 0.031).Conclusion In adults with Mallampati class 3, NTI using the BID caused an attenuated haemodynamic response and showed a higher success rate on the first attempt without increasing adverse events. The BID is an effective alternative to the Macintosh laryngoscope for NTI in anaesthetised adults with Mallampati class 3.
What problem does this paper attempt to address?