Research on Equivalent Circuit Model of Ultrasonic Bone Transducer

ZHANG Ganghua,CHEN Ying
DOI: https://doi.org/10.3969/j.issn.1004-2474.2011.04.019
2011-01-01
Abstract:Loss and tissue load model are two main factors that impact the precision of equivalent circuit model of ultrasonic bone transducer.Firstly,unloaded equivalent circuit model is developed considering piezoelectric loss,mechanical loss of piezoelectric stack and mechanical loss of other part of the bone transducer,which is verified by experiments that it can accurately predict unloaded properties of the transducer.Tissue load model and load related mechanical loss is given to describe properties of transducer with loads.Experiments show that this equivalent circuit model can well describe loading properties of ultrasonic bone transducer.
What problem does this paper attempt to address?